Automatically Determining the Confocal Parameters From OCT B-Scans for Quantification of the Attenuation Coefficients.

Dwork, Nicholas; Smith, Gennifer T; Leng, Theodore; Pauly, John M; Bowden, Audrey K · IEEE Trans Med Imaging · 2019

basic_science · Level V

Where this comes from

Abstract

The attenuation coefficient is a relevant biomarker for many diagnostic medical applications. Recently, the Depth-Resolved Confocal (DRC) technique was developed to automatically estimate the attenuation coefficients from Optical Coherence Tomography (OCT) data with pixel-level resolution. However, DRC requires that the confocal function parameters (i.e., focal plane location and apparent Rayleigh range) be known a priori. In this paper, we present the autoConfocal algorithm: a simple, automatic method for estimating those parameters directly from OCT imagery when the focal plane is within the sample. We present autoConfocal+DRC results on phantom data, ex-vivo biological tissue data, and in-vivo clinical data.