A Novel Technique for Transcatheter Aortic Valve Replacement in Pure Aortic Regurgitation.

Unbehaun, Axel; Klein, Christoph; Buz, Semih; Meyer, Alexander; Wamala, Isaac; Potapov, Evgenij; Schoenrath, Felix; Falk, Volkmar et al. · Ann Thorac Surg · 2019

case_report · Level V

Where this comes from

Abstract

Transcatheter aortic valve replacement (TAVR) in pure aortic regurgitation (AR) is challenging, because of the subsequent difficulty in anchoring the transcatheter valve in a noncalcified device landing zone (DLZ). Prestenting can help to prepare a stable DLZ for TAVR in pure AR and prevent valve migration. Here we report the first-in-human implantation of an uncovered stent into a noncalcified aortic valve as a prestenting strategy to prepare an easy DLZ for TAVR in pure AR. We consider this technique a useful novel tool to improve device success, at least as long as specific TAVR devices for pure AR are lacking.

Medical subject headings