EANM practical guidance on uncertainty analysis for molecular radiotherapy absorbed dose calculations.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 30218316.
- Also identified by DOI 10.1007/s00259-018-4136-7 and PMC identifier 6208822.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
A framework is proposed for modelling the uncertainty in the measurement processes constituting the dosimetry chain that are involved in internal absorbed dose calculations. The starting point is the basic model for absorbed dose in a site of interest as the product of the cumulated activity and a dose factor. In turn, the cumulated activity is given by the area under a time-activity curve derived from a time sequence of activity values. Each activity value is obtained in terms of a count rate, a calibration factor and a recovery coefficient (a correction for partial volume effects). The method to determine the recovery coefficient and the dose factor, both of which are dependent on the size of the volume of interest (VOI), are described. Consideration is given to propagating estimates of the quantities concerned and their associated uncertainties through the dosimetry chain to obtain an estimate of mean absorbed dose in the VOI and its associated uncertainty. This approach is demonstrated in a clinical example.
Medical subject headings
- Neoplasms
- Practice Guidelines as Topic
- Radiotherapy Planning, Computer-Assisted