Level-Set Segmentation-Based Respiratory Volume Estimation Using a Depth Camera.
Where this comes from
- Record sourced from PubMed, PMID 30235149.
- Also identified by DOI 10.1109/JBHI.2018.2870859 and PMC identifier 7309325.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
In this paper, a method is proposed to measure human respiratory volume using a depth camera. The level-set segmentation method, combined with spatial and temporal information, was used to measure respiratory volume accurately. The shape of the human chest wall was used as spatial information. As temporal information, the segmentation result from the previous frame in the time-aligned depth image was used. The results of the proposed method were verified using a ventilator. The proposed method was also compared with other level-set methods. The result showed that the mean tidal volume error of the proposed method was 8.41% compared to the actual tidal volume. This was calculated to have less error than with two other methods: the level-set method with spatial information (14.34%) and the level-set method with temporal information (10.93%). The difference between these methods of tidal volume error was statistically significant [Formula: see text]. The intra-class correlation coefficient (ICC) of the respiratory volume waveform measured by a ventilator and by the proposed method was 0.893 on an average, while the ICC between the ventilator and the other methods were 0.837 and 0.879 on an average.
Medical subject headings
- Image Interpretation, Computer-Assisted
- Imaging, Three-Dimensional
- Lung Volume Measurements
- Thoracic Wall