Can the radiopaque marker in surgical swabs scratch orthopaedic implant surfaces?
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 30269182.
- Also identified by DOI 10.1007/s00590-018-2309-5.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine whether the radiopaque marker strip, which is woven in surgical swabs, causes measureable wear on metal implants at pressures typically used to wipe off fluid from their surface. Finger pressure used to wipe a surface was measured and used as a reference pressure for further testing. A tribological wear rig was then used to analyse the wear caused on polished titanium plates by a cobalt chromium pin (the control test), the pin covered by a surgical swab and the pin covered by a radiopaque marker strip. It was found that the cotton part or the radiopaque marker of surgical swabs on polished medical grade titanium plates caused no significant wear. In contrast severe scratching was observed from the cobalt chromium pin on its own. To our knowledge, this is the first study in the literature analysing the wear caused by the surgical swabs and radiopaque strip on metal implants. The results suggest that surgical swabs are safe to use on metallic implants at pressures typical of a wiping motion.
Medical subject headings
- Fiducial Markers
- Prostheses and Implants
- Prosthesis Implantation