Patient Dose Comparison for Intraoperative Imaging Devices Used in Orthopaedic Lumbar Spinal Surgery.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 30280142.
- Also identified by DOI 10.5435/JAAOSGlobal-D-18-00030 and PMC identifier 6145549.
- Licence recorded as CC BY-NC-ND.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to determine the amount of radiation exposure from intraoperative imaging during two-level and four-level lumbar fusions. Five imaging systems were studied: multidetector CT (MDCT) scanner (CT A); two mobile CT units (CT B and CT C); a C-arm (D); and fluoroscopy (E). Metal oxide semiconductor field effect transistor dosimeters measured doses at 25 organ locations using an anthropomorphic phantom. A fat-equivalent phantom was used to simulate an obese body mass index (BMI). The effective dose (ED) for C-arm D was estimated using commercial software. The ED for others was computed from the measured mean organ doses. EDs for a normal BMI patient, receiving a four-level fusion, are as follows: CT A (12.00 ± 0.30 mSv), CT B (5.90 ± 0.25 mSv), CT C (2.35 ± 0.44 mSv), C-arm D (0.44 mSv), and fluoroscopy E (0.30 ± 0.3 mSv). The rankings are consistent across all three BMI values except CT C and fluoroscopy E, which peaked in the overweight size because of system limitations. The other machines' ED trended with patient BMI. The dose reduction protocols were confirmed according to the manufacturer's specifications. The results of this study emphasize the need for the appropriate selection of the imaging system, especially because the type of device could have a substantial effect on patient radiation risk.
Anatomy
- lumbar spine