Bone Marrow Biopsy Needle Type Affects Core Biopsy Specimen Length and Quality and Aspirate Hemodilution.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 30307478.
- Also identified by DOI 10.1093/ajcp/aqy126.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Bone marrow biopsies are essential for evaluating patients with suspected or confirmed hematopoietic disorders or malignancies, but little is known about how biopsy needle type affects biopsy length and/or quality. We sought to compare bone marrow biopsy quality in specimens obtained with two different needles. A retrospective analysis was performed on bone marrow specimens obtained with manual single-bevel (n = 114) or triple-bevel (n = 166) needles. The lengths of evaluable marrow, core quality, and aspirate quality were assessed by blinded hematopathologists. The triple-bevel needle produced 1.33-mm shorter lengths of evaluable marrow than the single-bevel needle and was five times less likely to produce a specimen rated as "adequate" and 4.2 times more likely to produce crush artifact. The triple-bevel needle was also 2.4 times more likely to produce hemodilute aspirates. Bone marrow biopsy needle type affects the length of evaluable marrow and quality of core and aspirate specimens.