Late Onset Valve Dislocation of the Edwards Intuity Rapid-Deployment Bioprosthesis.

Wacker, Max; Slottosch, Ingo; Scherner, Maximilian; Varghese, Sam; Wippermann, Jens · Ann Thorac Surg · 2019

case_report · Level V

Where this comes from

Abstract

Over the last decade, aortic valve replacement with rapid-deployment biologic bioprostheses has become a common alternative to the use of mechanical or biologic stented valves for high-risk patients. A 63-year-old patient underwent uncomplicated valve replacement with the Edwards Intuity valve (Edwards Lifesciences, Irvine, CA). Two months postoperatively, the patient had progressive dyspnea. Echocardiography showed a dislocated aortic valve reaching into the left ventricular outflow tract and impairing the anterior mitral valve leaflet. Both valves were replaced by a conventional stented bioprosthesis. This case report describes the first case of valve migration for the Edwards Intuity Elite rapid-deployment aortic valve and discusses possible explanations.