Late Onset Valve Dislocation of the Edwards Intuity Rapid-Deployment Bioprosthesis.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 30315805.
- Also identified by DOI 10.1016/j.athoracsur.2018.08.047.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Over the last decade, aortic valve replacement with rapid-deployment biologic bioprostheses has become a common alternative to the use of mechanical or biologic stented valves for high-risk patients. A 63-year-old patient underwent uncomplicated valve replacement with the Edwards Intuity valve (Edwards Lifesciences, Irvine, CA). Two months postoperatively, the patient had progressive dyspnea. Echocardiography showed a dislocated aortic valve reaching into the left ventricular outflow tract and impairing the anterior mitral valve leaflet. Both valves were replaced by a conventional stented bioprosthesis. This case report describes the first case of valve migration for the Edwards Intuity Elite rapid-deployment aortic valve and discusses possible explanations.
Medical subject headings
- Aortic Valve Stenosis
- Bioprosthesis
- Coronary Stenosis
- Echocardiography, Transesophageal
- Foreign-Body Migration
- Heart Valve Prosthesis Implantation