Modular femoral neck failure after revision of a total hip arthroplasty: a finite element analysis.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 30317469.
- Also identified by DOI 10.1007/s00590-018-2314-8.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The authors report on a case of modular femoral neck fracture which appeared 21 months after revision of acetabular component. The revision surgery was performed 8 years after the primary total hip arthroplasty due to aseptic loosening of the acetabular component. During acetabular revision, the primary implanted short (S, - 3.5 mm) femoral head was also exchanged with extra-long (XL, + 7.0 mm) femoral head fitting the modular femoral neck with a longer lever arm. Numerical analysis has shown that this has resulted in a 19.9% increase in tensile stress at the neck-stem coupling during normal walking cycle. This could result in microcrack initiation and propagation and finally lead to modular neck failure of the otherwise well-fixed stem. Surgeons should avoid excessive loading of the exchangeable neck (dual-modular) femoral stem designs as the stem-neck couplings are subject to corrosion and are not as reliable as monoblock stems.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Prosthesis Failure
- Reoperation
- Stress, Mechanical