Prospective evaluation of ultra-low-dose contrast-enhanced 100-kV abdominal computed tomography with tin filter: effect on radiation dose reduction and image quality with a third-generation dual-source CT system.

Leyendecker, Pierre; Faucher, Vanina; Labani, Aissam; Noblet, Vincent; Lefebvre, François; Magotteaux, Paul; Ohana, Mickaël; Roy, Catherine · Eur Radiol · 2019

prospective_cohort · Level II

Where this comes from

Abstract

To investigate the radiation dose exposure, image quality, and diagnostic performance of enhanced 100-kVp abdominopelvic single-energy CT protocol with tin filter (TF). Ninety-three consecutive patients referred for a single-phase enhanced abdominopelvic CT were prospectively included after informed consent. They underwent in addition to a standard protocol (SP) an acquisition with TF. Both examinations were performed on a third-generation dual-source CT system (DSCT), in single energy, using automatic tube current modulation, identical pitch, and identical level of iterative reconstruction. Radiation metrics were compared. Size-specific dose estimates (SSDE), contrast to noise ratio (CNR), and figure of merit (FOM) were calculated. Diagnostic confidence for the assessment of a predetermined list of abdominal lesions was rated by two independent readers. The mean dose of the TF protocol was significantly lower (CDTI 1.56 ± 0.43 mGy vs. 8.13 ± 3.32, p < 0.001; SSDE 9.94 ± 3.08 vs. 1.93 ± 0.39, p < 0.001), with an effective dose close to 1 mSv (1.14 mSv ± 0.34; p < 0.001). TF group exhibited non-significant lower liver CNR (2.76 vs. 3.03, p = 0.56) and was more dose efficient (FOM 10.6 vs. 2.49/mSv, p < 0.001) in comparison to SP. The mean diagnostic confidence for visceral, bone, and peritoneal tumors was equivalent between both groups. Enhanced 100-kVp abdominopelvic CT acquired after spectral shaping with tin filtration can achieve similar diagnostic performance and CNR compared to a standard CT protocol, while reducing the radiation dose by 81%. • 100-kVp spectral filtration enables enhanced abdominal CT with high-dose efficiency. • The radiation dose reaches the 1-mSv range. • Predetermined abdominopelvic lesions can be assessed without impairing on diagnostic confidence.

Medical subject headings