Ultrasonic measurement of optic nerve sheath diameter: a non-invasive surrogate approach for dynamic, real-time evaluation of intracranial pressure.
Where this comes from
- Record sourced from PubMed, PMID 30361274.
- Also identified by DOI 10.1136/bjophthalmol-2018-312934 and PMC identifier 6691934.
- Licence recorded as CC BY-NC.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The current study aimed to identify whether ultrasonographic measurements of optic nerve sheath diameter (ONSD) could dynamically and sensitively evaluate real-time intracranial pressure (ICP). ONSD measurements were performed approximately 5 min prior to and after a lumbar puncture (LP). A total of 84 patients (mean±SD age, 43.5±14.7 years; 41 (49%) men; 18 patients with elevated ICP) were included in the study. The Spearman correlation coefficients between the two observers were 0.779 and 0.703 in the transverse section and 0.751 and 0.788 in the vertical section for the left and right eyes, respectively. The median (IQR) change in ONSD (ΔONSD) and change in ICP (ΔICP) were 0.11 (0.05-0.21) mm and 30 (20-40) mmH<sub>2</sub>O, respectively, for all participants. With a reduction in cerebrospinal fluid pressure, 80 subjects (95%) showed an immediate drop in ONSD; the median (IQR) decreased from 4.13 (4.02-4.38) mm to 4.02 (3.90-4.23) mm (p<0.001). Significant correlations were found between ONSD and ICP before LPs (r=0.482, p<0.01) and between ΔONSD and ΔICP (r=0.451, p<0.01). Ultrasonic measurement of ONSD can reflect the relative real-time changes in ICP.
Medical subject headings
- Intracranial Pressure
- Optic Nerve
- Ultrasonography