Test-Retest Reproducibility of <sup>18</sup>F-FDG PET/CT Uptake in Cancer Patients Within a Qualified and Calibrated Local Network.
Where this comes from
- Record sourced from PubMed, PMID 30361381.
- Also identified by DOI 10.2967/jnumed.118.209544 and PMC identifier 6495239.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Calibration and reproducibility of quantitative <sup>18</sup>F-FDG PET measures are essential for adopting integral <sup>18</sup>F-FDG PET/CT biomarkers and response measures in multicenter clinical trials. We implemented a multicenter qualification process using National Institute of Standards and Technology-traceable reference sources for scanners and dose calibrators, and similar patient and imaging protocols. We then assessed SUV in patient test-retest studies. <b>Methods:</b> Five <sup>18</sup>F-FDG PET/CT scanners from 4 institutions (2 in a National Cancer Institute-designated Comprehensive Cancer Center, 3 in a community-based network) were qualified for study use. Patients were scanned twice within 15 d, on the same scanner (<i>n</i> = 10); different but same model scanners within an institution (<i>n</i> = 2); or different model scanners at different institutions (<i>n</i> = 11). SUV<sub>max</sub> was recorded for lesions, and SUV<sub>mean</sub> for normal liver uptake. Linear mixed models with random intercept were fitted to evaluate test-retest differences in multiple lesions per patient and to estimate the concordance correlation coefficient. Bland-Altman plots and repeatability coefficients were also produced. <b>Results:</b> In total, 162 lesions (82 bone, 80 soft tissue) were assessed in patients with breast cancer (<i>n</i> = 17) or other cancers (<i>n</i> = 6). Repeat scans within the same institution, using the same scanner or 2 scanners of the same model, had an average difference in SUV<sub>max</sub> of 8% (95% confidence interval, 6%-10%). For test-retest on different scanners at different sites, the average difference in lesion SUV<sub>max</sub> was 18% (95% confidence interval, 13%-24%). Normal liver uptake (SUV<sub>mean</sub>) showed an average difference of 5% (95% confidence interval, 3%-10%) for the same scanner model or institution and 6% (95% confidence interval, 3%-11%) for different scanners from different institutions. Protocol adherence was good; the median difference in injection-to-acquisition time was 2 min (range, 0-11 min). Test-retest SUV<sub>max</sub> variability was not explained by available information on protocol deviations or patient or lesion characteristics. <b>Conclusion:</b><sup>18</sup>F-FDG PET/CT scanner qualification and calibration can yield highly reproducible test-retest tumor SUV measurements. Our data support use of different qualified scanners of the same model for serial studies. Test-retest differences from different scanner models were greater; more resolution-dependent harmonization of scanner protocols and reconstruction algorithms may be capable of reducing these differences to values closer to same-scanner results.
Medical subject headings
- Fluorodeoxyglucose F18
- Neoplasms
- Positron Emission Tomography Computed Tomography