Pulmonary Cryoablation Zones: More Aggressive Ablation Is Warranted In Vivo.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 30383408.
- Also identified by DOI 10.2214/AJR.18.19527.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The objective of our study was to determine the effective cryoablation zone when treating pulmonary tumors in vivo and to create pulmonary-specific ablation maps to guide clinical procedure planning. Ablation volume was measured retrospectively in human patients after pulmonary tumor cryoablation with a triple-freeze protocol. Single-probe ablations were performed with 17-, 14-, and 13-gauge cryoprobes; multiple-probe ablations were performed with two or three 17-gauge probes. Statistical comparisons of ablation volumes to manufacturer reference values were calculated using the Wilcoxon rank-sum test. Comparisons of ablation sizes by the number of probes were evaluated by the Kruskal-Wallis test. Mean volume of in vivo lung ablation with a single 17-gauge cryoprobe measured 3.0 cm<sup>3</sup>, which is a statistically significant difference compared with the in vitro -20°C isotherm volume of 22.6 cm<sup>3</sup> (p < 0.01). Mean ablation volume of larger 13- and 14-gauge cryoprobes were 4.3 and 1.8 cm<sup>3</sup>, respectively, both of which are smaller than the in vitro -20°C isotherm volume. Mean cryoablation zone was not significantly affected by distance to the pleura (p = 0.54) or distance to a vessel (p = 0.55). Ablation volume was significantly increased (p < 0.01) with the use of multiple cryoprobes, at a rate of a 10.8-cm<sup>3</sup> increase per additional probe. The increased ablation zone size was more attributable to increased short-axis width (9.6-mm increase per probe) compared with long-axis length (5.6-mm increase per probe). The in vivo effective pulmonary cryoablation zone is significantly smaller than the manufacturer-published in vitro isotherm. Larger ablation margins in lung are best achieved by using multiple cryoprobes.
Medical subject headings
- Cryosurgery
- Lung Neoplasms