Heart Rate and Heart Rate Variability From Single-Channel Video and ICA Integration of Multiple Signals.
Where this comes from
- Record sourced from PubMed, PMID 30418892.
- Also identified by DOI 10.1109/JBHI.2018.2880097.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Unobtrusive monitoring of vital signs is relevant for both medical (patient monitoring) and non-medical applications (e.g., stress and fatigue monitoring). In this paper, we focus on the use of imaging photoplethysmography (iPPG). High frame rate videos were acquired by using a monochrome camera and an optical band-pass filter ([Formula: see text] nm). To enhance iPPG signal, we investigated the use of independent component analysis (ICA) pre-processing applied to iPPG signal from different regions of the face. Methodology was tested on [Formula: see text] healthy volunteers. Heart rate (HR) and standard time and frequency domain descriptors of heart rate variability (HRV), simultaneously extracted from videos and ECG data, were compared. A mean absolute error (MAE) about 3.812 ms was observed for normal-to-normal intervals with or without ICA pre-processing. Smaller MAE values of frequency domain descriptors were observed when ICA pre-processing was used. The impact of both video frame rate and video signal interval were also analyzed. All the results support the conclusion that proposed ICA pre-processing can effectively improve the HR and HRV assessment from iPPG.
Medical subject headings
- Heart Rate
- Photoplethysmography
- Signal Processing, Computer-Assisted
- Video Recording