Analytical and Clinical Performance Evaluation of the Elecsys HIV combi PT Assay on the cobas e 602 Analyzer for the Diagnosis of Human Immunodeficiency Virus.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 30423023.
- Also identified by DOI 10.1093/ajcp/aqy153 and PMC identifier 6396746.
- Licence recorded as CC BY-NC.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We evaluated the performance of the Elecsys HIV combi PT assay on the cobas e 602 analyzer for diagnosing human immunodeficiency virus (HIV; part of the US Food and Drug Administration [FDA] submission). The HIV combi PT and reference (ARCHITECT HIV Ag/Ab Combo) assays were assessed at four independent clinical laboratories/one reference laboratory (United States; July 2014 to November 2015). Clinical performance was evaluated using four reagent lots. Analytical performance was evaluated per Clinical and Laboratory Standards Institute EP05-A3 guidelines. Serum/plasma samples from 18 clinical sites/vendors (United States and outside the United States) were tested. Sensitivity (95% confidence interval [CI]) in HIV-1 antibody-positive individuals (United States and outside the United States; n = 1,460) was 100.00% (99.75%-100.00%). Specificity was 99.94% (95% CI, 99.85%-99.98%) in low-risk individuals (United States; n = 6,843), 98.19% (95% CI, 96.93%-99.04%) in high-risk individuals (United States and outside the United States; n = 758), and 97.43% (95% CI, 95.32%-98.76%) in pregnant women (United States and outside the United States; n = 440). Analytical performance was acceptable. We demonstrate the robustness of the FDA-approved Elecsys HIV combi PT assay on the cobas e 602 analyzer for HIV testing in the United States.
Medical subject headings
- HIV Antibodies
- HIV Infections
- HIV-1