High-yield paper-based quantitative blood separation system.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 30444230.
- Also identified by DOI 10.1039/c8lc00717a and PMC identifier 6287273.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Interest in developing paper-based devices for point-of-care diagnostics in resource-limited settings has risen remarkably in recent decades. In this paper, we demonstrate what we refer to as "high yield passive rrythrocyte removal" (HYPER) technology, which utilizes capillary forces in a unique cross-flow filtration for the separation of whole blood with performance comparable to centrifuges. As we will demonstrate, state-of-the-art passive blood separation methods implemented in paper-based systems exhibit rapid blood cell clogging on the filtration media or serum outlet and yield only about 10-30% of the total serum present in the sample. Our innovation results from the inclusion of a differentiation pad, which exploits hydrodynamic effects to reduce the formation of a fouling layer on the blood filtration membrane resulting in more than 60% serum yield with undiluted whole blood as direct input. To demonstrate the effectiveness of the HYPER technology we implement it in a lateral flow system and demonstrate the accurate quantification of vitamin A and iron levels in whole blood samples in 15 minutes.
Medical subject headings
- Blood Chemical Analysis
- Lab-On-A-Chip Devices