Prospective US investigational device exemption trial of a sutureless aortic bioprosthesis: One-year outcomes.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 30553598.
- Also identified by DOI 10.1016/j.jtcvs.2018.08.121.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We performed a prospective, single-arm clinical trial approved under a Food and Drug Administration Investigational Device Exemption to assess safety and efficacy of Perceval, a sutureless bovine pericardial aortic valve representing the initial US experience. From June 2013 to January 2015, 300 patients (mean age 76.7 ± 7.7 years, 54.3% men, 37.3% ≥80 years, median Society of Thoracic Surgeons Predicted Risk of Mortality 2.8%), underwent Perceval valve implantation at 18 centers across the United States. Twenty patients (6.7%) had a bicuspid aortic valve and 5 (1.7%) patients had previous aortic valve replacement. A minimally invasive approach was used in 80 (26.7%) and concomitant procedures were performed in 113 (37.8%) patients. Two hundred eighty-nine patients (96.3%) were successfully implanted. Operative mortality (≤30 days) was 1.3% (n = 4) (observed to expected ratio of 0.40). One-year results included all-cause mortality in 5.2% (n = 15), stroke in 1% (n = 3), and endocarditis in 1.7% (n = 5). New permanent periprocedural pacemaker rate was 10.7% (n = 30/281); 2.5% (n = 7/281) resulted from third-degree atrioventricular block. One-year valve-related reoperation was 2.1% (n = 6). At 1-year follow-up, 98% of patients were in New York Heart Association class I/II, left ventricular mass index decreased from 103.5 ± 30.1 g/m<sup>2</sup> at discharge to 95.8 ± 27.1 g/m<sup>2</sup> (P = .001), and 3 (1.3%) moderate paravalvular leaks were identified. Health-related quality of life score increased from 62.7 ± 21.8 before surgery to 85.5 ± 17.8 at 1 year (P < .001). These results confirm the safety and effectiveness of the Perceval sutureless aortic valve replacement in study patients with lower mortality than expected from a risk prediction model. Persistent hemodynamic benefit and improvement in quality of life at 1 year support the importance of this device in the management of aortic valve disease.
Medical subject headings
- Aortic Valve
- Bioprosthesis
- Heart Valve Diseases
- Heart Valve Prosthesis
- Heart Valve Prosthesis Implantation
- Sutureless Surgical Procedures