Early Distal Migration of a Self-Expanding Aortic Valve Prosthesis Causing Myocardial Infarction.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 30593781.
- Also identified by DOI 10.1016/j.athoracsur.2018.11.050.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Early distal migration after transcatheter aortic valve replacement is a rare but potentially catastrophic complication that presents unique technical challenges for subsequent surgical management. This report describes a case of early distal migration of a Medtronic CoreValve Evolut R bioprosthesis (Minneapolis, MN) causing myocardial infarction from coronary ostial obstruction and provides a practical technique for open surgical device explantation and aortic valve re-replacement. Snaring the stent of the device using standard instruments is a simple but effective method for transcatheter aortic valve replacement explant that allows for optimal positioning of a single aortotomy at the standard anatomic site to facilitate subsequent surgical aortic valve replacement.
Medical subject headings
- Bioprosthesis
- Myocardial Infarction
- Transcatheter Aortic Valve Replacement