Implementation and assessment of the black body bias correction in quantitative neutron imaging.
Where this comes from
- Record sourced from PubMed, PMID 30608985.
- Also identified by DOI 10.1371/journal.pone.0210300 and PMC identifier 6319815.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
We describe in this paper the experimental procedure, the data treatment and the quantification of the black body correction: an experimental approach to compensate for scattering and systematic biases in quantitative neutron imaging based on experimental data. The correction algorithm is based on two steps; estimation of the scattering component and correction using an enhanced normalization formula. The method incorporates correction terms into the image normalization procedure, which usually only includes open beam and dark current images (open beam correction). Our aim is to show its efficiency and reproducibility: we detail the data treatment procedures and quantitatively investigate the effect of the correction. Its implementation is included within the open source CT reconstruction software MuhRec. The performance of the proposed algorithm is demonstrated using simulated and experimental CT datasets acquired at the ICON and NEUTRA beamlines at the Paul Scherrer Institut.
Medical subject headings
- Image Processing, Computer-Assisted
- Neutron Diffraction
- Neutrons