Susceptibility of Cardiac Troponin Assays to Biotin Interference.
Where this comes from
- Record sourced from PubMed, PMID 30715102.
- Also identified by DOI 10.1093/ajcp/aqy172.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To investigate biotin interference on three cardiac troponin (cTn) assays and demonstrate a method to overcome biotin interference. cTn levels were measured in (1) plasma from healthy volunteers on 10-mg daily biotin supplementation mixed with a plasma with known elevated troponin, (2) plasmas with known elevated cTn after mixing in reagent biotin to simulate supplementation, and (3) biotin-spiked plasma specimens pretreated with streptavidin-agarose beads. Daily biotin ingestion (10 mg) and studies simulating daily biotin use resulted in significant interference in the Gen5 cardiac troponin T (cTnT) assay; the contemporary Gen 4 cTnT and high-sensitivity cardiac troponin I (hs-cTnI) assays were unaffected. The biotin interference threshold was 31, 315, and more than 2,000 ng/mL for Gen5 cTnT, cTnT, and hs-cTnI assays, respectively. Streptavidin pretreatment blocked biotin interference in cTn assays. Biotin interference is possible at plasma concentrations achievable by ingestion of over-the-counter supplements that may lead to delayed or missed diagnosis of myocardial injury with the Gen5 cTnT assay.
Medical subject headings
- Biotin
- Troponin T