Optimum Wavelengths in the Near Infrared for Imaging Photoplethysmography.
Where this comes from
- Record sourced from PubMed, PMID 30716029.
- Also identified by DOI 10.1109/TBME.2019.2897284.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this contribution is to determine the ideal near infrared wavelength bands for monochromatic and dual-band remote heartbeat detection using imaging photoplethysmography (iPPG) of the forehead. Experimental data of 38 healthy volunteers has been recorded and analyzed. For the data acquisition, a fast hyperspectral imager has been used. A new combination approach has been implemented that computes the quotient of the bands and, therefore, reduces motion artifacts. With this dual-band method excellent results (1.67 beats per minute mean deviation from electrocardiogram measurements for 73 recordings) have been obtained using a simple algorithm to analyze images at 799 and 861 nm. It can be concluded that excellent imaging photoplethysmography measurements can be performed at low cost using conventional silicon-based image sensors with invisible light in the near infrared region. This approach is a contribution to the development of non-contact heart rate measurement systems that can be used for medical diagnosis or other applications.
Medical subject headings
- Heart Rate
- Photoplethysmography