IGBT-Based Pulsed Electric Fields Generator for Disinfection: Design and In Vitro Studies on Pseudomonas aeruginosa.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 30726513.
- Also identified by DOI 10.1007/s10439-019-02225-0.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Irreversible electroporation of cell membrane with pulsed electric fields is an emerging physical method for disinfection that aims to reduce the doses and volumes of used antibiotics for wound healing. Here we report on the design of the IGBT-based pulsed electric field generator that enabled eradication of multidrug resistant Pseudomonas aeruginosa PAO1 on the gel. Using a concentric electric configuration we determined that the lower threshold of the electric field required to kill P. aeruginosa PAO1 was 89.28 ± 12.89 V mm<sup>-1</sup>, when 200 square pulses of 300 µs duration are delivered at 3 Hz. These parameters disinfected 38.14 ± 0.79 mm<sup>2</sup> area around the single needle electrode. This study provides a step towards the design of equipment required for multidrug-resistant bacteria disinfection in patients with pulsed electric fields.
Medical subject headings
- Disinfection
- Drug Resistance, Multiple, Bacterial
- Electromagnetic Fields
- Pseudomonas aeruginosa