A "one-stop-shop" 4D CTA protocol using 320-row CT for advanced imaging in acute ischemic stroke: a technical note.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 30770970.
- Also identified by DOI 10.1007/s00330-019-06041-x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This technical note describes a novel CT scan protocol that includes a non-enhanced CT, dynamic CTA, and perfusion of the whole brain and CTA of the carotid arteries using a 320-row area detector CT scanner, with a unique contrast injection and acceptable radiation exposure dose in patients presenting with acute ischemic stroke. The acquisition parameters and reconstruction parameters will be discussed including the use of model-based iterative reconstruction (MBIR), time summing (tMIP), and subtraction techniques to optimize the results of this protocol.Key Points• Scanning on a 320-row area detector CT can achieve both brain perfusion with dynamic angiography and reconstructed arterial and venous CTA, and supra aortic trunk angiography, in a single acquisition. • It provides, in a single exam, a full diagnostic workup, i.e., all the acquisitions that are needed to make a quick decision, with reasonable exposure to ionizing radiation and reduced amount of medium contrast, in case of acute ischemic stroke presentation.
Medical subject headings
- Brain Ischemia
- Computed Tomography Angiography
- Stroke