Audiology Telemedicine Evaluations: Potential Expanded Applications.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 30832542.
- Also identified by DOI 10.1177/0194599819835541 and PMC identifier 6602851.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
There is underutilization of cochlear implants with delays in implantation linked to distance from implant centers. Telemedicine could connect cochlear implant specialists with patients in rural locations. We piloted telemedicine cochlear implant testing in a small study, largely composed of normal-hearing volunteers to trial this new application of teleaudiology technology. Thirteen subjects (8 with normal hearing and 5 with hearing loss ranging from mild to profound) underwent a traditional cochlear implant evaluation in person and then via telemedicine technology. Routine audiometry, word recognition testing, and Arizona Biological Test (AzBio) and consonant-nucleus-consonant (CNC) testing were performed. Mean (SD) percent difference in AzBio between in-person and remote testing was 1.7% (2.06%). Pure tone average (PTA), speech reception threshold (SRT), and word recognition were similar between methods. CNC testing showed a mean (SD) difference of 6.8% (10.2%) between methods. Testing conditions were acceptable to audiologists and subjects. Further study to validate this method in cochlear implant candidates and a larger population is warranted.
Medical subject headings
- Audiology
- Cochlear Implantation
- Cochlear Implants