Early Detection of Hypoglycemia in Type 1 Diabetes Using Heart Rate Variability Measured by a Wearable Device.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 30877089.
- Also identified by DOI 10.2337/dc18-1843.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Changes in heart rate variability (HRV) occur at the initiation of hypoglycemia due to sympathetic nervous system activity. We investigated the use of HRV detection by a wearable device as an early alert for hypoglycemia. This proof-of-principle study included 23 patients with type 1 diabetes (14 women, mean age 42 ± 11 years). Patients wore a VitalConnect HealthPatch for 5 days. Hypoglycemia was defined as glucose ≤70 mg/dL (≤3.9 mmol/L) by fingerstick measurement. HRV was analyzed in standardized periods before the hypoglycemic event. Sixty-six hypoglycemic events were recorded. Hypoglycemia caused a typical HRV pattern in 36 (55%) of the hypoglycemic events. Eighteen events (27%) showed an atypical pattern. Ten events were unclassified (15%), and two did not display a change in HRV (3%). Hypoglycemia causes early changes in HRV that can be detected by a wearable device. Measuring real-time HRV seems promising for early hypoglycemia detection.
Medical subject headings
- Diabetes Mellitus, Type 1
- Hypoglycemia
- Wearable Electronic Devices