Reduction of measurement noise in a continuous glucose monitor by coating the sensor with a zwitterionic polymer.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 30931173.
- Also identified by DOI 10.1038/s41551-018-0273-3 and PMC identifier 6436621.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Continuous glucose monitors (CGMs), used by patients with diabetes mellitus, can autonomously track fluctuations in blood glucose over time. However, the signal produced by CGMs during the initial recording period following sensor implantation contains substantial noise, requiring frequent recalibration via fingerprick tests. Here, we show that coating the sensor with a zwitterionic polymer, found via a combinatorial-chemistry approach, significantly reduces signal noise and improves CGM performance. We evaluated the polymer-coated sensors in mice as well as in healthy and diabetic non-human primates, and show that the sensors accurately record glucose levels without the need for recalibration. We also show that the polymer-coated sensors significantly abrogated immune responses to the sensor, as indicated by histology, fluorescent whole-body imaging of inflammation-associated protease activity, and gene expression of inflammation markers. The polymer coating may allow CGMs to become standalone measuring devices.
Medical subject headings
- Biosensing Techniques
- Blood Glucose
- Coated Materials, Biocompatible
- Polymers