The risk of loosening of extramedullary fracture fixation devices.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 30955874.
- Also identified by DOI 10.1016/j.injury.2019.03.051.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Extramedullary devices that use screws, pins or wires are used extensively to treat fractures in normal and diseased bone. A common failure mode is implant loosening at the bone-screw/pin/wire interface before fracture healing occurs. This review first considers the fundamental mechanics of the bone-fixator construct with focus on interfacial strains that result in loosening. It then evaluates the time-independent and time-dependent material models of bone that have been used to simulate and predict loosening. It is shown that the recently developed time-dependent models are capable of predicting loosening due to cyclic loads in bone of varying quality.
Medical subject headings
- Fracture Fixation
- Fracture Healing
- Fractures, Bone