The effect of vitamin E-enhanced cross-linked polyethylene on wear in shoulder arthroplasty-a wear simulator study.

Alexander, Justin J; Bell, Simon N; Coghlan, Jennifer; Lerf, Reto; Dallmann, Frank · J Shoulder Elbow Surg · 2019

biomechanical · Level V

Where this comes from

Abstract

Wear of the polyethylene glenoid component and subsequent particle-induced osteolysis remains one of the most important modes of failure of total shoulder arthroplasty. Vitamin E is added to polyethylene to act as an antioxidant to stabilize free radicals that exist as a byproduct of irradiation used to induce cross-linking. This study was performed to assess the in vitro performance of vitamin E-enhanced polyethylene compared with conventional polyethylene in a shoulder simulator model. Vitamin E-enhanced, highly cross-linked glenoid components were compared with conventional ultrahigh-molecular-weight polyethylene glenoids, both articulating with a ceramic humeral head component using a shoulder joint simulator over 500,000 cycles. Unaged and artificially aged comparisons were performed. Volumetric wear was assessed by gravimetric measurement, and wear particle analysis was also subsequently performed. Vitamin E-enhanced polyethylene glenoid components were found to have significantly reduced wear rates compared with conventional polyethylene in both unaged (36% reduction) and artificially aged (49% reduction) comparisons. There were no differences detected in wear particle analysis between the 2 groups. Vitamin E-enhanced polyethylene demonstrates improved wear compared with conventional polyethylene in both unaged and artificially aged comparisons and may have clinically relevant benefits.