Efficacy and Usability of a Novel Barrier Device for Preventing Injection Port Contamination: A Pilot Simulation Study.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 31136328.
- Also identified by DOI 10.1213/ANE.0000000000004235.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Contamination of intravenous (IV) ports and stopcocks has been associated with postoperative infections. We tested the usability and efficacy of a novel passive shielding device to prevent such contamination even in the absence of hand hygiene or port disinfection. In a desktop setting with deliberately contaminated hands, qualitative port contamination was detected after 5/60 (8.3%; 95% confidence interval [CI], 2.8-18.4) control port injections versus 0/60 (0%; 95% CI, 0-6.0) shielded injections (P = .025). In clinical simulations with a quantitative bioburden assay (measured in relative light units [RLUs]), median (interquartile range [IQR]) postsimulation bioburden was 46 (32-53) vs 27 (21-42) RLU for the control versus intervention groups (P = .036), yielding a median shift of -13 RLU (95% CI, -2 to -26) in favor of the shielding. Usability of the device was acceptable to practitioners.