An Innovative Approach to Airway Foreign Body Management in an Extremely Premature Neonate.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 31241769.
- Also identified by DOI 10.1002/lary.28153.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This case report is a retrospective review of a challenging though ultimately successful removal of an airway foreign body in a 2-day-old premature female born at 23 weeks and 4 days gestation. A segment of a 5-F surfactant catheter was cut and accidentally dislodged in the distal airway within the lumen of the patient's endotracheal tube. Ultimately, visualization was obtained using a 1.6-mm KARL STORZ sialoendoscope, and retrieval was performed using sialoendoscopy forceps passed via the working channel of the sialoendoscope while maintaining ventilation. This case represents an innovative approach to complex airway foreign body management utilizing sialoendoscopy instruments. Laryngoscope, 130:1236-1238, 2020.
Medical subject headings
- Bronchi
- Foreign Bodies
- Intubation, Intratracheal