Microfluidic protein isolation and sample preparation for high-resolution cryo-EM.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 31292253.
- Also identified by DOI 10.1073/pnas.1907214116 and PMC identifier 6660723.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
High-resolution structural information is essential to understand protein function. Protein-structure determination needs a considerable amount of protein, which can be challenging to produce, often involving harsh and lengthy procedures. In contrast, the several thousand to a few million protein particles required for structure determination by cryogenic electron microscopy (cryo-EM) can be provided by miniaturized systems. Here, we present a microfluidic method for the rapid isolation of a target protein and its direct preparation for cryo-EM. Less than 1 μL of cell lysate is required as starting material to solve the atomic structure of the untagged, endogenous human 20S proteasome. Our work paves the way for high-throughput structure determination of proteins from minimal amounts of cell lysate and opens more opportunities for the isolation of sensitive, endogenous protein complexes.
Medical subject headings
- Cryoelectron Microscopy
- Image Processing, Computer-Assisted
- Proteasome Endopeptidase Complex
- Protein Subunits