Abstract
Artificial smokes focusing on macroscopic or fluid properties of smoke have been available for a long time. This paper presents a simple method to generate fully customizable smoke-like atmospheres at microscopic scale (i.e. considering their constituent particles as discrete elements) using a different approach. Synthetic, reproducible media can be generated combining monodisperse microspheres with known geometrical and optical properties conveniently parameterized. The method is presented as a proof-of-concept, highlighting the design decisions along with their implications. Practical issues such as aerosol nebulization, particle carrier selection or the features of the medium chamber where the smoke-like atmosphere is to be tested are analyzed. A comparison between methanol and ethanol as carriers for polystyrene microsphere nebulization is also made. The method could be the seed for the obtention of standard reference media for calibration or standardized characterization of not only smoke detectors and exhaust smoke sensors but also other instruments relying on optical properties of dispersive media (dust in PV panels, public lighting, etc.).