A needle-to-post air discharge ion source in tandem with FAIMS system.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 31419259.
- Also identified by DOI 10.1371/journal.pone.0221080 and PMC identifier 6697363.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A needle-to-post ionization source was designed for high-field asymmetric waveform ion mobility spectrometry (FAIMS). The needle-to-post ion source includes asymmetric electrode comprised of a copper post with a diameter of 2 mm and a stainless-steel needle with 200-μm tip radius and length of 28 mm. With the discharge voltage of -5.6 kV and N2 gas flow, glow discharge was realized at atmospheric pressure. The mass spectra of ionized ions about acetone, ethanol and ethyl acetate were gotten by Thermo Scientific LTQ XL ion trap mass spectrometer (MS). The MS experimental results show that the main ions are protonated and dimer ions. The needle-to-post ion source was mounted on the FAIMS system and FAIMS spectra are gotten successfully. Separation of p-xylene, o-xylene and m-xylene was realized. It shows that the needle-to-post electrode could be used as the ion source in a FAIMS system.
Medical subject headings
- Electrodes
- Ion Mobility Spectrometry
- Xylenes