Micro-dislodgement during transcatheter aortic valve implantation with a contemporary self-expandable prosthesis.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 31697759.
- Also identified by DOI 10.1371/journal.pone.0224815 and PMC identifier 6837455.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the incidence, risk factors and the clinical outcome of micro-dislodgement (MD) with a contemporary self-expandable prosthesis during transcatheter aortic valve implantation. MD was defined as movement of the prosthesis of at least 1.5 mm upwards or downwards from its position directly before release compared to its final position. Patients were grouped according to the occurrence (+MD) or absence (-MD) of MD. Baseline characteristics, imaging data and outcome parameters were retrospectively analyzed. We identified 258 eligible patients. MD occurred in 31.8% (n = 82) of cases with a mean magnitude of 2.8 mm ± 2.2 in relation to the left coronary cusp and 3.0 mm ± 2.1 to the non-coronary cusp. Clinical and hemodynamic outcomes were similar in both groups with consistency over a follow-up period of three months. A larger aortic valve area (AVA) (-MD vs. +MD: 0.6 cm2 ± 0.3 vs. 0.7cm2 ± 0.2; p = 0.014), was the only independent risk factor for the occurrence of MD in a multivariate regression analysis (OR 5.3; 95% CI: 1.1-24.9; p = 0.036). MD occurred in nearly one third of patients and did not affect clinical and hemodynamic outcome. A larger AVA seems to be a potential risk factor for MD.
Medical subject headings
- Heart Valve Prosthesis
- Prosthesis Failure
- Self Expandable Metallic Stents
- Transcatheter Aortic Valve Replacement