Evaluation of an AI-based, automatic coronary artery calcium scoring software.
Level III
Where this comes from
- Record sourced from PubMed, PMID 31728692.
- Also identified by DOI 10.1007/s00330-019-06489-x and PMC identifier 7033052.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
To evaluate an artificial intelligence (AI)-based, automatic coronary artery calcium (CAC) scoring software, using a semi-automatic software as a reference. This observational study included 315 consecutive, non-contrast-enhanced calcium scoring computed tomography (CSCT) scans. A semi-automatic and an automatic software obtained the Agatston score (AS), the volume score (VS), the mass score (MS), and the number of calcified coronary lesions. Semi-automatic and automatic analysis time were registered, including a manual double-check of the automatic results. Statistical analyses were Spearman's rank correlation coefficient (⍴), intra-class correlation (ICC), Bland Altman plots, weighted kappa analysis (κ), and Wilcoxon signed-rank test. The correlation and agreement for the AS, VS, and MS were ⍴ = 0.935, 0.932, 0.934 (p < 0.001), and ICC = 0.996, 0.996, 0.991, respectively (p < 0.001). The correlation and agreement for the number of calcified lesions were ⍴ = 0.903 and ICC = 0.977 (p < 0.001), respectively. The Bland Altman mean difference and 1.96 SD upper and lower limits of agreements for the AS, VS, and MS were - 8.2 (- 115.1 to 98.2), - 7.4 (- 93.9 to 79.1), and - 3.8 (- 33.6 to 25.9), respectively. Agreement in risk category assignment was 89.5% and κ = 0.919 (p < 0.001). The median time for the semi-automatic and automatic method was 59 s (IQR 35-100) and 36 s (IQR 29-49), respectively (p < 0.001). There was an excellent correlation and agreement between the automatic software and the semi-automatic software for three CAC scores and the number of calcified lesions. Risk category classification was accurate but showing an overestimation bias tendency. Also, the automatic method was less time-demanding. • Coronary artery calcium (CAC) scoring is an excellent candidate for artificial intelligence (AI) development in a clinical setting. • An AI-based, automatic software obtained CAC scores with excellent correlation and agreement compared with a conventional method but was less time-consuming.
Medical subject headings
- Artificial Intelligence
- Calcium
- Coronary Angiography
- Coronary Artery Disease
- Coronary Vessels
- Software
- Tomography, X-Ray Computed