Validation of a custom spine biomechanics simulator: A case for standardization.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 31740014.
- Also identified by DOI 10.1016/j.jbiomech.2019.109470 and PMC identifier 6952059.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Mechanical testing machines used in cadaveric spine biomechanics research vary between labs. It is a necessary first step to understand the capabilities and limitations in any testing machine prior to publishing experimental data. In this study, a reproducible protocol that uses a synthetic spine was developed and used to quantify the inherent rotation error and the ability to apply loads in a single physiologic plane (pure-moment) of a custom spine biomechanics simulator. Rotation error was evaluated by comparing data collected by the test machine and the data collected by an optical motion capture system. Pure-moment loading was assessed by comparing the out-of-plane loads to the primary plane load. Using synthetic functional spine units previously shown to have mechanics similar to the cadaveric human spine, the simulator was evaluated using a dynamic test protocol reflective of its future use in the study of cadaveric spine specimens. Rotation errors inherent in the test machine were <0.25° compared to motion capture. Out of plane loads were <4.0% of the primary plane load, which confirmed pure-moment loading. The authors suggest that a standard validation protocol for biomechanical spine testing machines is needed for transparency and accurate field-wide data interpretation and comparison. We offer recommendations based on the reproducible use of a synthetic spinal specimen for consideration.
Medical subject headings
- Materials Testing
- Mechanical Phenomena
- Spine