Safety of Noncontrast Imaging-Guided Deep Brain Stimulation Electrode Placement in Parkinson Disease.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 31756502.
- Also identified by DOI 10.1016/j.wneu.2019.11.071.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Deep brain stimulation (DBS) is considered standard of care for the treatment of medically refractory Parkinson disease (PD). The placement of brain electrodes is performed using contrast imaging to enhance blood vessel identification during stereotactic planning. We present our experience with a series of patients implanted using noncontrast imaging. All cases of DBS surgery for PD performed between 2012 and 2018 with noncontrast imaging were retrospectively reviewed. Clinical features, postoperative imaging, and complications were analyzed. A total of 287 deep-seated electrodes were implanted in 152 patients. Leads were placed at the subthalamic nucleus and globus pallidus internus in 258 and 29 hemispheres, respectively. We identified 2 cases of intracranial hemorrhage (0.7%). DBS lead placement can be performed without the use of intravenous contrast with a postoperative intracranial hemorrhage rate comparable with other reported series.
Medical subject headings
- Implantable Neurostimulators
- Intracranial Hemorrhages
- Magnetic Resonance Imaging
- Neurosurgical Procedures
- Parkinson Disease
- Postoperative Hemorrhage
- Prosthesis Implantation
- Surgery, Computer-Assisted