AUTOMATED RETINAL LAYER SEGMENTATION AND THEIR THICKNESS PROFILES IN HEALTHY SUBJECTS: A Comparison of 55° Wide-field and Conventional 30° Spectral Domain-Optical Coherence Tomography.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 31834134.
- Also identified by DOI 10.1097/IAE.0000000000002714 and PMC identifier 7505155.
- Licence recorded as CC BY-NC-ND.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To assess whether retinal thickness measurements with a standard 30° spectral domain optical coherence tomography (SD-OCT) are comparable with wide-field 55° SD-OCT. Thirty-three healthy individuals were scanned using 55° as well as 30° SD-OCT according to a standardized protocol. Automated retinal layer segmentation of standard and wide-field SD-OCTs was assessed using customized software. Both lenses showed a high correlation when analyzing total retinal thickness within the central, the inner, and the outer retinal ring (r = > 0.9). Automated thickness measurements with the 55° system were marginally higher compared with the 30° lens. The thickness of each separate retinal layer using automated segmentation showed excellent correlations within the inner and outer rings (range: r = 0.6-r = 0.9 for the inner ring and range: r = 0.9-r = 1.0 for the outer ring). Fifty-five degree wide-field SD-OCT provides a good overview of the posterior pole and presents similar quantitative values as a standard 30° OCT lens. Therefore, thickness values are comparable when switching between these two lenses.
Medical subject headings
- Retina
- Tomography, Optical Coherence