Microscope-Induced Thermal Burns During Lymphaticovenular Anastomosis.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 31850967.
- Also identified by DOI 10.1097/SAP.0000000000002116.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Bright illumination sources using xenon lamps have improved microsurgical visualization under an operating microscope; however, surgeons must recognize the potential for accidental thermal damage to soft tissues.In this article, we present 2 reports of microscopic thermal burn in lymphaticovenular anastomosis (LVA).A 23-year-old woman and a 57-year-old woman with bilateral lymphedema of the legs had LVAs on both legs under local anesthesia. The burn wound in a 23-year-old woman was full thickness, and the one in a 57-year-old woman was deep dermal burn. Both of them healed without skin grafting.Working distance and high illumination intensity are important risk factor. The use of epinephrine as part the local anesthetic mixture that decreases blood flow is also a major risk factor for thermal burns. Lymphaticovenular anastomosis particularly requires high magnification, which leads to increasing the intensity and decreasing the working distance. The surgical conditions around LVA are inherently prone to microscope-induced thermal burns.
Medical subject headings
- Burns
- Lymphatic Vessels
- Lymphedema