MRI in patients with implanted active devices: how to combine safety and image quality using a limited transmission field?
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 31974690.
- Also identified by DOI 10.1007/s00330-019-06599-6.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Radio frequency (RF) pulses in magnetic resonance imaging (MRI) can interact with implanted devices and cause tissue damage. However, there are new devices that can safely perform measurements with liberal MRI conditions such as an RF transmission field B<sub>1+rms</sub> ≤ 2.0 μT. We investigated whether MRI in this case is limited for these technical reasons. We selected typical MRI protocols of six body regions (brain, cervical spine, lumbar spine, knee, liver, heart) using two typical 1.5T MRI scanners. Overall, we adapted 62 sequences to B<sub>1+rms</sub> conditions and evaluated their diagnostic quality. For this, we measured signal-to-noise-ratio (SNR), contrast-to-noise-ratio (CNR), and geometric deviation (GD) as quality parameters, using phantom studies. For questionnaire studies, we selected pairs of original and adapted sequences in healthy volunteers. Blinded radiologists rated the images as single sequence rating and in direct comparison. Roughly one-third of the checked sequences were below the B<sub>1+rms</sub> limit. Here, 56 of the 62 adapted sequences showed at least the same image quality in single ratings. A reduction in SNR and/or CNR was found with 31 sequences and only one sequence with considerably increased GD. Especially, sequences with original high B<sub>1+rms</sub> values, PD sequences, and sequences of the Siemens knee and heart protocol were difficult to adapt, whereas most TSE and IR sequences had no clinical limitations. By limiting the transmission field to B<sub>1+rms</sub> ≤ 2.0 μT, clinically relevant MR sequences can be adapted with nearly no reduction in image quality. Despite limiting the transmission field, high-quality MR imaging is possible. We could derive strategies for adaptation. • Despite limiting the transmission field, high-quality MRI is possible. • We could derive strategies for adapting the sequences to B<sub>1+rms</sub>≤ 2.0 μT. • This enables high-quality MRI of different body regions for patients with AD.
Medical subject headings
- Brain
- Liver
- Lumbar Vertebrae
- Magnetic Resonance Imaging
- Phantoms, Imaging
- Prostheses and Implants
- Radio Waves