False positive results of Bowie and Dick type test used for hospital steam sterilizer with slower come-up ramps: A case study.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 31986177.
- Also identified by DOI 10.1371/journal.pone.0227943 and PMC identifier 6984693.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine if the standardized B&D type test for hospital steam sterilizer is correctly indicating cycle failures in slower come-up ramps cycles. Two commercially available B&D type test were challenged in a quality control sterilizer. A common failure was simulated in triplicate cycles, using a standardized cycle configuration, and then compared to triplicate cycles of a common cycle configuration. Tests procedures were conducted according to B&D manufacturer test standard and results were compared to standardized endpoint specifications. We found that B&D type tests are only capable of detecting the presence of non-condensable gases if the sterilization equipment is adjusted to meet all the cycle requirements of the test. False positive results were obtained with come-up ramp time of 3 min. Correct results were only obtained with come-up ramp of 1.7-1.9 min. Until the ISO 17665 and AAMI ST-79 standards are revised, equipment qualification experts should observe come-up time duration criterion for B&D type test cycles according to ISO 11140-4. Sterile Processing Department professionals must add the come-up ramp criterion to cycle evaluation before clearing the equipment for routine use. This will allow B&D correct performance, reducing the infection risk from unsterilized medical device.
Medical subject headings
- Equipment and Supplies, Hospital
- Quality Control
- Steam
- Sterilization