Acquisition time, radiation dose, subjective and objective image quality of dual-source CT scanners in acute pulmonary embolism: a comparative study.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 32025830.
- Also identified by DOI 10.1007/s00330-019-06650-6.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To compare the scan acquisition time, radiation dose, subjective and objective image quality of two dual-source CT scanners (DSCT) for detection of acute pulmonary embolism. Two hundred twenty-one scans performed on the 2nd-generation DSCT and 354 scans on the 3rd-generation DSCT were included in this large retrospective study. In a randomized blinded design, two radiologists independently reviewed the scans using a 5-point Likert scale. Radiation dose and objective image quality parameters were calculated. Mean acquisition time was significantly lower in the 3rd-generation DSCT (2.81 s ± 0.1 in comparison with 9.7 s ± 0.15 [mean ± SD] respectively; p < 0.0001) with the 3rd generation 3.4 times faster. The mean subjective image quality score was 4.33/5 and 4/5 for the 3rd- and 2nd-generation DSCT respectively (p < 0.0001) with strong interobserver reliability agreement. DLP, CTDI<sub>vol</sub>, and ED were significantly lower in the 3rd than the 2nd generation (175.6 ± 63.7 mGy cm; 5.3 ± 1.9 mGy and 2.8 ± 1.2 mSv in comparison with 266 ± 255 mGy.cm; 7.8 ± 2.2 mGy and 3.8 ± 4.3 mSv). Noise was significantly lower in the 3rd generation (p < 0.01). Signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and figure of merit (FOM), a dose-insensitive index for CNR, were significantly higher in the 3rd-generation DSCT (33.5 ± 23.4; 29.0 ± 21.3 and 543.7 ± 1037 in comparison with 23.4 ± 17.7; 19.4 ± 16.0 and 170.5 ± 284.3). Objective and subjective image quality are significantly higher on the 3rd-generation DSCT with significantly lower mean acquisition time and radiation dose. • The 3rd-generation DSCT scanner provides an improved image quality, less perceived artifacts, and lower radiation dose in comparison with the 2nd-generation DSCT, when operating in dual-energy (DE) mode. • The 3.4-times-faster 3rd-generation DSCT scanner can be of particular value in patients with chronic lung diseases or breathing difficulties that prevent adequate breathhold.
Medical subject headings
- Computed Tomography Angiography
- Pulmonary Embolism
- Radiation Dosage
- Tomography Scanners, X-Ray Computed