Dose assessment for patients with stage I non-small cell lung cancer receiving passive scattering carbon-ion radiotherapy using daily computed tomographic images: A prospective study.

Li, Yang; Kubota, Yoshiki; Kubo, Nobuteru; Mizukami, Tatsuji; Sakai, Makoto; Kawamura, Hidemasa; Irie, Daisuke; Okano, Naoko et al. · Radiother Oncol · 2020

prospective_cohort · Level II

Where this comes from

Abstract

This study aimed to assess dose distributions for stage I non-small cell lung cancer (NSCLC) with passive scattering carbon-ion radiotherapy (C-ion RT) using daily computed tomography (CT) images. We enrolled 10 patients with stage I NSCLC and acquired a total of 40 pre-fractional CT image series under the same settings as the planning CT images. These CT images were registered with planning CT images for dose evaluation using both bone matching (BM) and tumor matching (TM). Using deformable image registration, we generated accumulated doses. Moreover, the volumetric dose parameters were compared in terms of tumor coverage and lung exposure and statistical analyses were performed. Overall, 25% of 40 fractional dose distributions were unacceptable with BM, compared with 2.5% with TM (P < 0.001). Using BM, three patients' accumulated dose distributions were unacceptable; however, all were satisfactory with TM (P < 0.001). No differences were observed in water-equivalent path length (WEL). The required margins in patients with poor dose distribution were 5.9 and 4.4 mm for BM and TM, respectively. This study establishes that CT image-based TM is robust compared with conventional BM for both daily and accumulated dose distributions. The effects of changes in WEL seem to be limited. Hence, daily CT alignment is recommended for patients with stage I NSCLC receiving C-ion RT.

Medical subject headings