Evaluation of bacterial proliferation with a microfluidic-based device: Antibiochip.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 32107504.
- Also identified by DOI 10.1371/journal.pone.0223932 and PMC identifier 7046301.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The measurement of the proliferation (and the relevant inhibition of proliferation) of microbes is used in different settings, from industry to laboratory medicine. Thus, in this study, the capacity of the Antibiochip (ELTEK spa), a microfluidic-based device, to measure the amount of E. coli in certain culture conditions, was evaluated. An Antibiochip is composed of V-shaped microchannels, and the amount of microparticles (such as microbes) is measured by the surface of the pellet after centrifugation. In the present study, different geometries, volumes and times were analyzed. When the best conditions were identified, serial dilutions of microbial cultures were tested to validate the linearity of the results. Then, with the use of wild E. coli strains isolated from medical samples, the relationship between bacterial susceptibility to antibiotics measured by standard methods and that measured by the Antibiochip was evaluated. In this report, the good quality performances of the methods, their linearity and the capacity to identify susceptible microbial strains after 60 minutes of incubation are shown. These results represent a novel approach for ultrarapid antibiograms in clinics.
Medical subject headings
- Bacteria
- Cell Proliferation
- Lab-On-A-Chip Devices