Application of Attachable Magnetic Nerve Stimulator in Intraoperative Facial Nerve Monitoring during Ear Surgery.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 32151184.
- Also identified by DOI 10.1177/0194599820907912.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We developed an attachable magnetic nerve stimulator (AMNS) that connects the metallic instruments to a neurophysiological monitoring unit for monitoring the facial nerve (FN) during ear surgery and present our experiences with intraoperative neuromonitoring (IONM) of the FN using AMNS. The FN in 20 patients who underwent tympanomastoidectomy for chronic ear disease was examined. The electromyography (EMG) amplitudes of the FN using AMNS were assessed. The EMG amplitudes of the FN at 1.00-mA stimulation during drilling were 265 ± 64 µV in the orbicularis oris and 288 ± 57 µV in the orbicularis oculi. The EMG amplitudes using AMNS attached to the surgical instruments under 0.35-mA stimulus at the tympanic segment of the FN were 196 ± 43 µV in the orbicularis oris and 197 ± 41 µV in the orbicularis oculi. The application of continuous stimulation with burr and surgical instruments using the AMNS is feasible and effective for FN stimulation and identification.
Medical subject headings
- Facial Nerve
- Otologic Surgical Procedures