Mechanical Effects of Cochlear Implants on Residual Hearing Loss: A Finite Element Analysis.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 32191879.
- Also identified by DOI 10.1109/TBME.2020.2981863.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The effects of cochlear implants on residual hearing loss is investigated through a finite element model of human auditory periphery consisting of the cochlea and middle ear. The simulation results show that a round window stiffness is the dominant factor in residual hearing loss. The increased round window stiffness to five times caused over 4 dB residual hearing loss at low frequencies below 500 Hz. Without considering round window ossification, inserting a cochlear implant can show at most 4 dB difference of residual hearing loss in magnitude from the no-implant case although the cochlear implant's geometry and position has been varied. If the stiffness of the round window is the same, the simulation results suggest to use a thin-straight-cochlear implant inserted into the lateral side in order to preserve residual hearing at frequencies below 700 Hz. In addition, when the distance between the basilar membrane and a cochlear implant is closer, the residual hearing loss becomes severe at high frequencies above 1 kHz. The results would be helpful for choice of a cochlear implant depending on a patient's condition.
Medical subject headings
- Cochlear Implantation
- Cochlear Implants
- Hearing Loss