Drilling Energy Correlates With Screw Insertion Torque, Screw Compression, and Pullout Strength: A Cadaver Study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 32251146.
- Also identified by DOI 10.5435/JAAOS-D-19-00366.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine whether drilling energy correlates with bone mineral density (BMD), maximum insertion torque (MIT), maximum screw compression, and pullout strength (POS). Ten cadaver tibias were used for testing. Unicortical pilot holes were drilled and the drilling energy measured. Drill site bone quality was determined with microcomputed tomography. Drill holes were randomly assigned to POS or MIT testing using 3.5-mm cortical screws engaging only the near cortex. Pearson correlation coefficients were calculated to determine the relationship between drilling energy, BMD, POS, MIT, and maximum screw compression. Drilling energy was correlated with BMD (P < 0.001). Compared with BMD, drilling energy had a better correlation with MIT, maximum screw compression, and POS. Maximum screw compression also correlated with MIT (P = 0.012). Drilling energy better correlates with MIT, maximum screw compression, and POS compared with BMD in cadaver cortical bone. Dynamically measuring drilling energy may help inform the orthopaedic surgeon as to the quality of the bone before insertion of implants.
Medical subject headings
- Biomechanical Phenomena
- Bone Density
- Bone Screws
- Mandrillus
- Orthopedic Procedures
- Tibia
- Torque