Near-infrared fluorescence imaging for intraoperative margin assessment during robot-assisted partial nephrectomy.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 32320124.
- Also identified by DOI 10.1111/bju.15089.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To demonstrate how using a standardised dosing strategy of indocyanine green (ICG) dye with near-infrared fluorescence (NIRF) imaging can be used to differentiate renal tumours from normal renal parenchyma during robot-assisted partial nephrectomy (RAPN). We reviewed a prospectively collected database of 361 consecutive RAPNs using NIRF between June 2011 and March 2018, and determined the rate at which differential fluorescence was achieved. Tumour and kidney fluorescence or afluorescence were recorded intraoperatively and compared to histological results on final pathology. Of 330 tumours, after 31 exclusions for nonvisible tumours due to adherent fat, completely intrarenal location or for incomplete data, 288 (87.3%) successfully exhibited differential fluorescence. Among the predominant histologies, 249 of 277 (89.9%) renal cell carcinomas did not fluoresce, as well as 23 of 32 (71.9%) oncocytomas. Real-time gross assessment of resection margin for fluorescence yielded a positive margin rate on final pathology of 0.30%. When administered with a standardised ICG-dosing strategy, NIRF successfully achieved differential fluorescence in a large majority of tumours during RAPN with an exceedingly low positive margin rate.
Medical subject headings
- Coloring Agents
- Indocyanine Green
- Kidney Neoplasms
- Margins of Excision
- Nephrectomy
- Optical Imaging
- Robotic Surgical Procedures