Bifurcation Aneurysm Treatment with PulseRider: Technical Nuances, Device Selection, Positioning Strategies, and Overcoming Geometric Challenges.
Level V
Where this comes from
- Record sourced from PubMed, PMID 32360671.
- Also identified by DOI 10.1016/j.wneu.2020.04.126.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Endovascular therapy is a viable alternative to surgical clipping for the treatment of intracranial aneurysms; however, aneurysms arising at bifurcations remain a challenge. The purpose of this technical report is to share the nuances of treating aneurysms with the PulseRider (PR), including device selection and positioning strategy, from authors who are highly experienced in its use. We offer a comprehensive guide for neuroendovascular surgeons less experienced with PR applications to include design, general coil embolization technique, principles of deployment and detachment, positioning options, and geometric challenges and their solutions. In our experience, the PR is well suited for addressing the challenges of treating bifurcation aneurysms. PR use is intuitive and straightforward for use in bifurcation aneurysms with ideal favorable. PR can also be safely and effectively used to address a much broader and more challenging range of geometries.
Medical subject headings
- Blood Vessel Prosthesis
- Embolization, Therapeutic
- Endovascular Procedures
- Intracranial Aneurysm
- Prosthesis Design
- Stents