Plasma Isolation in a Syringe by Conformal Integration of Inertial Microfluidics.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 32367467.
- Also identified by DOI 10.1007/s10439-020-02526-9.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A thermoplastic microfluidic substrate is conformally integrated onto the cylindrical barrel of a conventional venipuncture syringe, forming a spiral inertial separation element supporting the isolation of plasma from diluted whole blood. The cylindrical shape of the syringe itself serves to define the flow path required for inertial separation by transforming a linear microchannel to a spiral topology. The hybrid system enables inertial plasma separation by Dean flow focusing within the same syringe used for a patient blood draw, with the seamlessly interconnected microfluidic element operated by automated or manual actuation of the syringe plunger. Plasma isolation is achieved without the need for external instrumentation. Device design and fabrication challenges are discussed, and effective plasma isolation within the system is demonstrated, with a peak separation efficiency above 97% using 25 × diluted blood.
Medical subject headings
- Cell Separation
- Microfluidics
- Plasma
- Syringes